Z-2668-2025
device
Siemens Medical Solutions USA, Inc
MAGNETOM Vida Fit. Model Number: 11410481.
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the ...
Ongoing
null
2025-08-28
2025-10-08
null
41 units
null
2240869
3
0
Voluntary Action Indicated (VAI)
2022-10-26
0
0
null
false
0
device_malfunction
venting system blockage
Z-0459-2026
device
Abbott Medical
Liberta RC DBS IPG (Implantable Pulse Generator), Model Number 62400
There is a potential issue affecting Eterna Spinal Cord Stimulation (SCS) Implantable Pulse Generators (IPGs), Model 32400 and Liberta RC Deep Brain Stimulation (DBS) Implantable Pulse Generators (IPGs), Model 62400 . These devicess may experience a loss of communication with the Clinician Programmer (CP) and/or the Pa...
Ongoing
null
2025-10-06
2025-11-19
null
722 units
null
3009600098
3
0
No Action Indicated (NAI)
2014-01-29
0
0
null
false
0
device_malfunction
communication loss
Z-0116-2020
device
GE Healthcare, LLC
Aespire 7100, Model Numbers: 1. 1009-9000-000-009469 2. 1009-9000-000 3. 1009-9000-000-013913 4. 1009-9000-000-001366 5. 1009-9000-000-009857 6. 1009-9003-000 7. 1009-9000-000-008120 8. 1009-9000-000-009600 9. 1009-9000-000-014290 10. 1009-9000-000-010585 11. 1009-9000-000-004993 12. 1009-9011-000-004795 ...
Certain Aespire and Aestiva Anesthesia Systems were noted to have a vulnerability to a cyber-attack when connected to the hospital network. An insufficiently secured terminal server may provide an opportunity for a malicious actor that has already penetrated the hospital network to send fraudulent flow sensor correcti...
Ongoing
null
2019-07-19
2019-12-25
null
19351 devices
null
null
null
null
null
null
0
0
null
false
null
software
cybersecurity vulnerability
Z-1563-2025
device
Baxter Healthcare Corporation
Welch Allyn Welch Allyn Connex Spot Monitor, Product Code/ Part Numbers: 1) 73WT-B; 2) 74CE-B; 3) 74CT-B; 4) 74CX-B; 5) 74ME-B; 6) 74MT-B; 7) 74MX-B; 8) 74RE-B; 9) 74RT-B; 10) 75CE-B; 11) 75CT-B; 12) 75CT-BR; 13) 75CX-B; 14) 75-HCA-CTB; 15) 75-HCA-MTB; 16) 75ME-B; 17) 75MT-B; 18) 75...
Undeclared natural rubber band within the packaging of certain configurations of reusable blood pressure cuffs.
Ongoing
null
2025-03-21
2025-04-23
null
70627 units
null
1025114
11
5
Official Action Indicated (OAI)
2024-02-09
0
0
null
true
0
foreign_material
rubber band contamination
Z-0697-2025
device
MEDLINE INDUSTRIES, LP - Northfield
HUDSON RCI mBrace, ET Tube Holder with 4 Point Head Strap, REF DYNJMBRC4; tracheal tube fixation device
Complaint defects have been noted regarding the adhesive elements of the device and the frame of the holder putting additional pressure on the patient's cheekbones and upper lip which may cause dermal injury depending on the device positioning.
Ongoing
null
2024-11-18
2024-12-25
null
1464 units
null
null
null
null
null
null
0
0
null
false
null
device_malfunction
adhesive defect
D-0475-2021
drug
Cardinal Health Inc.
Prenatal Tablets Gluten Free Multivitamin/ Multimineral Dietary Supplement for Pregnant and Lactating Women UD 100 Tablets (10x10) NDC 77333-715-10 GenDose Pharmaceuticals
CGMP Deviations: Intermittent exposure to temperature excursion during storage.
Terminated
null
2021-03-15
2021-06-02
FL, GA, SC
41 bottles
null
3003474118
3
0
No Action Indicated (NAI)
2025-02-19
0
0
null
false
-1
cGMP
temperature excursion
Z-3399-2018
device
B. Braun Medical, Inc.
EC20O 20G X 100CM CATH OPEN TIP LF, Material Number 333520 Connection device used by physicians to provide various anesthetic and fluid administration devices with a single, common access point to a catheter for delivery of anesthetics. The connector is used in conjunction with the catheter for continuous administr...
The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the catheters from the catheter connectors has been observed.
Terminated
null
2018-07-20
2018-10-10
null
3900
null
2523676
11
0
Voluntary Action Indicated (VAI)
2025-05-16
0
0
null
false
-1
device_malfunction
connector seal failure
D-1312-2019
drug
Inopak Ltd
Instant Waterless Hand Sanitizer, Ethyl Alcohol, 62%, packaged as a) DermaGel Instant Waterless Hand Sanitizing Gel with Moisturizers, with Aloe Vera & Vitamin E, packaged in a) 1000 ml Disc Pump pouches (NDC 58575-340), 8 x 1000 ml Disc Pump pouches per case, Product Code 5025-L1000; b) 2 fl. oz. bottles, 80/2 oz bott...
CGMP Deviations: hand sanitizers and soaps were not produced under current good manufacturing practices.
Terminated
null
2019-04-22
2019-06-05
Nationwide in the USA
17,850 pounds per batch
null
1000122232
7
3
Official Action Indicated (OAI)
2023-08-03
2
2
2023-12-15
false
-1
cGMP
manufacturing practices violation
D-0066-2017
drug
Pharmedium Services, LLC
lidocaine HCl, 2%, Preservative Free, Total Volume 2mL, Product code 2R3323, NDC 61553-323-79
Stability Data Does Not Support Expiry: potential loss of potency in drugs packaged and stored in syringes.
Terminated
null
2015-07-17
2016-10-19
Nationwide
830 syringes
null
3000717703
4
3
Official Action Indicated (OAI)
2017-12-22
1
1
2014-07-18
false
-1
stability
potency loss syringes
D-1299-2022
drug
Eco Lips, Inc
Juice Beauty, The Organic Solution, SPF 8, Naturally Clear Lip Moisturizer, Hydratant pour les Levres, NET WT 0.15 oz (4.25 g) per tube, Active Ingredient: Zinc oxide 4%, Manufactured for Juice Beauty, Inc., 709 Fifth Ave., San Rafael, California 94901-3566
Failed Dissolution Specifications
Terminated
null
2022-06-29
2022-08-03
CA
35,313 tubes
null
null
null
null
null
null
0
0
null
false
null
potency
failed dissolution
Z-0133-2019
device
Zimmer Biomet, Inc.
Affixus Hip Fracture Nail Right 125 11 mm x 360 mm, Item Number 814311360 Product Usage: Intended for the fixation of fractures
There were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Terminated
null
2018-08-08
2018-10-24
null
12
null
null
null
null
null
null
0
0
null
false
null
sterility
compromised packaging
Z-0827-2022
device
Mevion Medical Systems, Inc.
Mevion S250 and S250i Proton Therapy Systems hand pendant (P/N 10263 or P/N 37808)
Malfunction of the hand pendant controlling movement of the therapy couch and nozzle may result in unintentional motion. Patient injury is possible if they come in contact with the nozzle or other objects in the treatment room.
Ongoing
null
2022-02-01
2022-04-06
null
39 (36 US, 3 OUS)
null
3007087027
4
0
Voluntary Action Indicated (VAI)
2020-03-05
0
0
null
false
0
device_malfunction
control pendant failure
D-0379-2022
drug
Dr. Reddy's Laboratories, Inc.
Rugby, Antihistamine and Nasal Decongestant, Fexofenadine HCl 60mg and Pseudoephedrine HCl 120mg, Extended-Release Tablets USP, 30 Tablets per box, NDC 0536-1242-07, Distributed by: Rugby Laboratories, 17177 N Laurel Park Drive, Suite 233, Livonia, MI, 49152, Made in India.
Failed Dissolution Specifications
Terminated
null
2022-01-04
2022-01-26
Nationwide in the USA
2,352 boxes
null
3011009040
1
0
No Action Indicated (NAI)
2025-08-21
0
0
null
false
-1
potency
failed dissolution
D-1148-2017
drug
Bella Pharmaceuticals, Inc.
Mannitol 20%, 10 mL vials, Rx only, Bella Pharma 3101 W. Devon Ave., Chicago, IL 60659, 1 (877) BELLA.
Lack of Assurance of Sterility.
Terminated
null
2017-08-15
2017-09-20
Nationwide within USA
10 vials
null
3013442632
1
1
Official Action Indicated (OAI)
2017-07-18
1
1
2019-02-11
true
0
sterility
sterility assurance failure
Z-0594-2018
device
DePuy Orthopaedics, Inc.
CORAIL HIGH OFFSET STEM Collarless SIZE 14
Incorrect device in the package. A package that was labeled as the CORAIL COXA VARA HIGH OFFSET STEM Collared Size 9 (Part No. 3L93709 / Lot No. 5291990), contained the CORAIL HIGH OFFSET STEM Collarless SIZE 14 (Part No. L20314 / Lot No. 5292130).
Terminated
null
2017-07-12
2018-02-21
null
19
null
3005572635
1
0
No Action Indicated (NAI)
2009-01-12
0
0
null
false
0
mislabeling
incorrect device packaged
Z-1654-2014
device
Siemens Healthcare Diagnostics, Inc
ADVIA Centaur Systems TSH3 Ultra , Ready Pack, 500 test Catalog Number: 06491080; Siemens Material Number: 10282379. Intended for the invitro diagnostic use in quantitative determination of various assays on the ADVIA Centaur and ADVIA Centaur XP analyzers.
Samples containing fluorescein may show interference with the ADVIA Centaur¿ Systems TSH3 Ultra, Vitamin D and the ADVIA Centaur BRAHMS Procalcitonin assays.
Terminated
null
2014-04-10
2014-06-04
null
155,345 kits
null
null
null
null
null
null
0
0
null
false
null
software
assay interference
Z-2301-2014
device
Zimmer, Inc.
Persona (TM) The Personalized Knee System CPS Tibial Articular Surface Provisional Top Nonsterile TASP MLC L 3-5 CD TOP TASP MLC L 6-9 CD TOP TASP MLC L 3-5 EF TOP TASP MLC L 6-9 EF TOP TASP MLC L 10-11 EF TOP TASP MLC L 6-9 GH TOP
Zimmer had received complaints that reported breakage of the Persona Knee Trial Articular Surface Provisionals (TASPs). This is a retrospective recall which covers Zimmer's corrective actions taken earlier this year concerning the TASPs.
Terminated
null
2013-06-10
2014-08-27
null
70,986 distribution events
null
1000220733
10
4
No Action Indicated (NAI)
2022-07-27
0
0
null
false
-1
device_malfunction
breakage of prosthetic component
Z-1672-2023
device
Smith & Nephew, Inc.
Smith&Nephew ENGAGE Cementless Partial Knee System, Coated Tibial Tray, Part Numbers: a) REF 1-10012-100, SIZE 1-LEFT MEDIAL; b) REF 1-10012-150, SIZE 1-RIGHT MEDIAL; c) REF 1-10012-200, SIZE 2-LEFT MEDIAL; d) REF 1-10012-250, SIZE 2-RIGHT MEDIAL; e) REF 1-10012-300, SIZE 3-LEFT MEDIAL; f) REF 1-10012-350, S...
Recent complaint data indicates that the revision rate may be trending higher than corresponding similar devices in global joint replacement registries. The data identifies a potential signal that the performance is an outlier versus the state of the art with respect to the risk for revision.
Ongoing
null
2023-05-02
2023-06-07
null
2146 units
null
1020279
6
0
Voluntary Action Indicated (VAI)
2021-12-09
0
0
null
false
0
device_malfunction
higher revision rate
Z-3350-2018
device
B. Braun Medical, Inc.
CE18T CONT EPIDURAL SET, Material Number 332200 Connection device used by physicians to provide various anesthetic and fluid administration devices with a single, common access point to a catheter for delivery of anesthetics. The connector is used in conjunction with the catheter for continuous administration of an...
The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the catheters from the catheter connectors has been observed.
Terminated
null
2018-07-20
2018-10-10
null
10500
null
3014301563
1
0
Voluntary Action Indicated (VAI)
2018-12-27
0
0
null
false
-1
device_malfunction
connector failure
Z-1374-2024
device
Boston Scientific Corporation
Obsidio Conformable Embolic, Material Numbers (UPN): a) M0013972001010, b) M0013972101010
An investigation determined that delivery of the Obsidio embolic using the aliquot technique for lower gastrointestinal bleeding embolization poses a high risk of bowel ischemia. The most serious and the most common adverse health consequence, reasonably foreseeable to occur, is the need to perform major surgery such ...
Ongoing
null
2024-02-20
2024-04-03
null
985 units
null
3005099803
7
0
Voluntary Action Indicated (VAI)
2024-11-26
0
0
null
false
-1
device_malfunction
ischemia risk
Z-0716-2024
device
Olympus Corporation of the Americas
Soltive SuperPulsed Laser System TFL Fiber Stripper, Model Numbers TFL-AFS150, TFL-AFS200, TFL-AFS365, TFL-AFS550, TFL-AFS940
The strippers and cleavers are provided non-sterile and labelled as reusable/autoclavable. Olympus does not have validated cleaning and sterilization instructions. Use of a non-sterile fiber stripper or cleaver on a sterile fiber poses a risk of contamination.
Ongoing
null
2023-12-04
2024-01-24
null
3001 units
null
1000635022
4
0
Voluntary Action Indicated (VAI)
2022-03-10
1
1
2015-08-12
false
0
sterility
non-sterile device
Z-1452-2022
device
Stryker Instruments Div. of Stryker Corporation
SurgiCount+ System Application, Software Version 2.0.17, Catalog Number 0694-002-090. Used for counting, displaying, recording and locating the number of RFID-tagged absorbent articles used during surgical procedures.
When scanning sponges out after a surgical procedure, an error may allow for a duplicate sponge to be counted without the prompt to make a case note and allowing the case to close. The software update will prevent the occurrence of this error.
Terminated
null
2022-06-15
2022-07-27
null
44 units
null
null
null
null
null
null
0
0
null
false
null
software
sponge counting logic
D-1197-2015
drug
Lincare, Inc.
Vancomycin 1000mg / NS 100mL, Packaged in 100 mL Bags, Rx Only. Compounded by: United Medical IV Center (Little Rock) 1527 South Bowman Rd. Ste. D Little Rock, AR 72211-4200.
Lack of Assurance of Sterility: Sterility of product is not assured.
Terminated
null
2015-05-28
2015-07-08
United States including: Arkansas
Unknown
null
1000307060
1
0
Voluntary Action Indicated (VAI)
2010-09-22
0
0
null
false
0
sterility
sterility assurance failure
Z-1689-2025
device
Boston Scientific Corporation
IceSeed 1.5 CX NEEDLE OUS, Cryoablation Needle, REF H7493967534170. The needles are intended to convert high-pressure gas to either a very cold Freezing application or to a warm Thawing application.
Boston Scientific has identified that IceSeedTM CX needles manufactured between April 2024 and February 2025 have been programmed with DEMO settings instead of Commercial settings (see Affected Product Listing). Upon initial connection of an IceSeed CX needle to the Cryoablation System, the user will be prompted to per...
Ongoing
null
2025-03-24
2025-05-07
null
67 units
null
3002095335
11
0
No Action Indicated (NAI)
2025-03-18
0
0
null
false
0
software
incorrect demo settings
Z-0376-2022
device
Cardinal Health
RoyalSilk Non-Reinforced Surgical Gowns X-large SKU: 9548
Potential for open packaging seals compromising the sterility of the surgical gowns
Ongoing
null
2021-11-02
2021-12-22
null
270,839 units
null
3012150451
1
0
No Action Indicated (NAI)
2021-10-25
0
0
null
false
0
sterility
compromised package seals
Z-2483-2024
device
GE Medical Systems, LLC
SIGNA 7.0T, with affected software versions: 7T29.1, MR30.1; Nuclear Magnetic Resonance Imaging System
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR seque...
Ongoing
null
2024-06-14
2024-08-14
null
3 units
null
1000116747
1
0
No Action Indicated (NAI)
2016-10-11
0
0
null
false
0
device_malfunction
SAR limit exceed
Z-1847-2013
device
Hospira Inc.
The Hospira Symbiq Infusion System is intended for the delivery of fluids, solutions, drugs, agents, nutritionals, electrolytes, blood and blood products via parenteral, enteral, intravenous, intra-arterial, subcutaneous, epidural, or irrigation routes of administration.
The devices secondary audio signal fails to activate.
Terminated
null
2013-02-04
2013-08-07
null
2,791 units
null
null
null
null
null
null
0
0
null
false
null
device_malfunction
audio signal failure
D-0539-2016
drug
Walgreens Infusion Services
Ceftriaxone 2g/NS 100 mL Minibag Plus, Rx Only, Compounded by Option Care 148, San Antonio, TX 78232-5045
Lack of Assurance of Sterility
Terminated
null
2015-11-10
2016-01-06
Texas
46 Minibags
null
null
null
null
null
null
0
0
null
false
null
sterility
lack of sterility assurance
Z-1816-2025
device
Baxter Healthcare Corporation
Affinity 4 Birthing Bed, Product Codes P3700E100429, P3700E100432, P3700E100433, P3700E100434, P3700E100435, P3700E100436, P3700E100437, P3700E100438, P3700E100439, P3700E100441, P3700E100442, P3700E100443, P3700E100444, P3700E100445, P3700E100446, P3700E100448, P3700E100449, P3700E100450, P3700E100453, P3700E100455, P...
There is a potential for improper crimps on the signal wires of the actuators which could result in a loss of ability to lower the bed or the foot section of the bed.
Ongoing
null
2025-05-06
2025-06-04
null
164 units
null
1417572
11
1
No Action Indicated (NAI)
2024-02-06
1
1
2009-09-10
false
0
device_malfunction
improper wire crimp
Z-0994-2020
device
Cardinal Health 200, LLC
Genius 2 Tympanic Thermometer Private Label Item Code: 3069
The frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range (¿ 0.2¿C for Genius 2 and ¿ 0.3¿C for Genius 3 thermometers). The measurement readings drift upwards over time, which means that the thermome...
Terminated
null
2019-10-28
2020-02-12
null
in total 37567 units US; 335573 OUS
null
2016452
3
0
No Action Indicated (NAI)
2021-12-21
0
0
null
false
-1
device_malfunction
calibration drift accuracy
Z-2442-2018
device
Intel-GE Care Innovations LLC
Health Harmony Mobile application software Product Usage: Care Innovations Health Harmony Mobile is intended as a communication tool to display medical device data from third party devices for patients to view, and to collect assessment (question & answers) from patients in the home.Care Innovations Health Har...
It was discovered that in certain situations, including partial sessions and when taking adhoc measurements, the patient data was not synchronizing in a timely manner with the backend database, resulting in the patient's clinician not getting patient data tor one or two days.
Terminated
null
2017-12-06
2018-07-25
null
1302
null
3008846825
1
0
No Action Indicated (NAI)
2011-10-21
0
0
null
false
0
software
data sync delay
D-0884-2021
drug
Washington Homeopathic Products, Inc.
Z-1070 Formula Anxioheal. Contents 1000ml 20% Alcohol. WHP Homeopathic Medicine, For Manufacturing Use Only, REMEDY:Causticum 30X, Conium Maculatum 8X, Conium Maculatum 12X, Crataegus Oxyacantha 12X, Crataegus Oxyacantha 30X, Ignatia Amara 24X, Lycopus Virginicus 12X, Phosphoricum Acidum 12X, Phosphoricum Acidum 24X, R...
Out-of-specification test results obtained for identification testing based on HPTLC analysis.
Terminated
null
2021-08-20
2021-10-06
Distributed Nationwide in the USA
2/1 Liter bottle
null
3006489197
10
4
Voluntary Action Indicated (VAI)
2023-08-23
3
3
2020-06-19
false
-1
potency
failed identification testing
Z-2298-2018
device
Howmedica Osteonics Corp.
LFIT Anatomic CoCr Femoral Heads, Catalog No. 6260-9-336
The recalling firm has received a higher than expected number of complaints documenting femoral head/hip stem disassociation for certain sizes of femoral heads manufactured prior to March 4, 2011.
Terminated
null
2018-05-22
2018-07-11
null
null
3003070421
7
0
Voluntary Action Indicated (VAI)
2023-12-01
0
0
null
false
-1
device_malfunction
femoral head disassociation
Z-2079-2014
device
Siemens Healthcare Diagnostics Inc
RAPIDPoint 500 Measurement Clinical Chemistry System RAPIDPoint 500 Measurement Cartridge Lactate (750 Tests) Part number: 10491449; RAPIDPoint 500 Measurement Cartridge Lactate (400 Tests) Part Number: 10491448 RAPIDPoint 500 Measurement Cartridge Lactate (250 Tests) Part Number: 10491447 RAPIDPoint 500 Measure...
RAPIDPoint 500 Measurement Cartridges may have an error code one or more electrolyte parameters (e.g., Na+, Ca++, K+, Cl-)
Terminated
null
2014-05-23
2014-07-30
null
20,507 cartridges
null
3002637618
3
0
No Action Indicated (NAI)
2021-08-19
0
0
null
false
-1
device_malfunction
electrolyte measurement error
Z-0426-2016
device
bioMerieux, Inc.
MYLA MASTER DVD 3.2 CLI DL380 Product Usage: MYLA is a computer application (Middleware) based on Web 2.0 technology which: Interfaces between the instruments connected to the application and the LIS(s) (Laboratory Information System(s)).
MYLA¿ software connected to a Laboratory Information System (LIS) and a VITEK¿ 2 system has the potential to link a test result to an incorrect patient record with the same specimen ID.
Terminated
null
2015-11-03
2015-12-23
null
626 systems in total
null
3002769706
4
2
No Action Indicated (NAI)
2017-04-10
1
1
2012-08-23
false
-1
software
patient record linkage
Z-1819-2025
device
GE Medical Systems, LLC
OEC Elite: The OEC Elite mobile C-arm is intended to provide fluoroscopic and digital spot images of the patient anatomy, interventional tools/devices, and contrast agents during diagnostic, interventional, and surgical procedures.
OEC Elite and OEC 3D Mobile C-Arms Exceeding 4% Source to Image Distance in X-ray Field.
Ongoing
null
2025-03-21
2025-06-18
null
Total for both products: 3472 (2,919 US; 553 OUS)
null
1000116747
1
0
No Action Indicated (NAI)
2016-10-11
0
0
null
false
0
device_malfunction
X-ray field calibration
D-0030-2024
drug
Central Admixture Pharmacy Services Inc
HYDROmorphone in dextrose 5%, 3 mg/30 mL, (0.1 mg/mL), 30 mL Syringe, Rx only, Central Admixture Pharmacy Services, Inc., 7935 Dunbrook Road, Suite C, San Diego, CA 92126, NDC: 71286-2013-2
Lack of assurance of sterility:
Terminated
null
2023-09-21
2023-10-11
Nationwide in the USA
274 30 mL syringes
null
3005292119
1
0
Voluntary Action Indicated (VAI)
2021-05-24
0
0
null
false
0
sterility
sterility assurance failure
Z-0670-2013
device
Terumo Cardiovascular Systems Corporation
Sarns high-flow aortic arch cannula, 6.5 mm (20 Fr) OD with 3/8in connector, 11in (28 cm) long The Sarns" High Flow Aortic Arch Cannula and Sarns" Flexible Arterial Cannulae are indicated for use in perfusion of the asc...
Flexible Arterial Cannula and High-Flow Aortic Arch Cannula connectors were identified to contain flash on the exterior of the cannula connector that exceeded established specifications. Because the flash has the potential to detach in the operative field, TCVS has decided to remove all affected lots that were manufact...
Terminated
null
2013-01-07
2013-01-23
null
6830 total all products
null
3000204839
6
0
No Action Indicated (NAI)
2024-02-16
0
0
null
false
-1
device_malfunction
flash defect connector
D-0718-2023
drug
Central Admixture Pharmacy Services, Inc.
vancomycin added to 0.9% sodium chloride, 750 mg/250 mL* (3 mg/mL), 250 mL per bag, Rx only, Central Admixture Pharmacy Services, Inc., 6580 Snowdrift Rd., Ste 100, Allentown, PA 18106, NDC 71285-6071-1.
Lack of Assurance of Sterility: after an FDA inspection called into question the sterility of the products intended to be sterile.
Terminated
null
2023-04-28
2023-05-31
Nationwide in the USA.
3075 bags
null
3009590582
4
3
Official Action Indicated (OAI)
2023-03-30
3
3
2024-03-29
true
0
sterility
sterility assurance failure
Z-1854-2017
device
Smith & Nephew, Inc.
Peri-Loc(R) VLP 3.5 MM X 6 MM SELF-TAPPING CORTEX SCREW, QTY: (1), STERILE R, REF 71801306 Orthopedic
The expiration dates on the labeling are incorrect, the actual expiration dates are before the date on the labeling.
Terminated
null
2017-03-09
2017-04-26
null
1 unit
null
1219602
3
0
No Action Indicated (NAI)
2017-05-25
0
0
null
false
-1
mislabeling
incorrect expiration date
Z-1739-2024
device
Cardinal Health 200, LLC
Jackson-Pratt 3-Spring Reservoir with Silicone Adapters, 400mL, REF SU130-475
Cardinal Health is issuing a lot number specific voluntary removal of Jackson-Pratt 3-Spring Reservoir kits due to specific products being shipped to users before undergoing sterilization.
Ongoing
null
2024-03-27
2024-05-08
null
98 cases of 12
null
3002973487
3
0
No Action Indicated (NAI)
2015-07-20
0
0
null
false
0
sterility
non-sterilized product
Z-2599-2023
device
Stryker Corporation
Stryker ACM (Advanced Cement Mixing) System: ACM Kit w/ Femoral Breakaway Nozzle & Solid Blades-Allows the user to mix bone cement with a constant high vacuum through the process Catalog Number: 0306-564-000
Potential for the Break-Away Femoral Nozzle the applicator tip connecting the cement cartridge, to disassemble or break off during use, can lead to the potential for loss of function of the ACM system and bio-incompatibility due to components falling into surgical site
Ongoing
null
2023-08-07
2023-09-27
null
8 units US
null
3015967359
1
0
Voluntary Action Indicated (VAI)
2023-01-24
0
0
null
false
0
device_malfunction
femoral nozzle breakage
Z-1816-2021
device
Alphatec Spine, Inc.
Atec Insignia Anterior Cervical Plate System, REF 136-0350 Insignia, ACP, 3-Level, 50 mm, Rx Only, Non-Sterile, UDI: (01)00190376268569
There is a potential for the screw blocking mechanism on the anterior cervical plate system to disassociate intraoperatively/postoperatively or unlock postoperatively.
Terminated
null
2021-04-26
2021-06-16
null
N/A
null
2027467
8
2
No Action Indicated (NAI)
2022-07-18
3
3
2015-07-16
false
-1
device_malfunction
screw mechanism failure
Z-0915-2022
device
Boston Scientific Corporation
ROTAWIRE Drive and wireClip Torquer, ROTAWIRE Drive Floppy
Pouch seals may be open, compromising sterility.
Ongoing
null
2022-03-08
2022-04-20
null
433 5-packs
null
3003882783
1
0
No Action Indicated (NAI)
2018-08-30
0
0
null
false
0
sterility
open pouch seals
D-0636-2019
drug
King Bio Inc.
Dr. King's Natural Medicine Homeopathic Ear Ringing, 2 fl. oz. (59 mL), King Bio, Inc., 3 Westside Dr. Asheville, NC 28806, UPC 357955527527, NDC 57955-5275-2
Microbial contamination
Terminated
null
2018-07-20
2019-03-20
U.S.A. Nationwide, Canada, and Australia.
1980 bottles
null
1053442
8
1
Voluntary Action Indicated (VAI)
2023-07-28
2
2
2019-03-20
false
-1
contamination
microbial contamination
Z-3386-2018
device
B. Braun Medical, Inc.
GOVCE17TKFC EPIDURAL TRAY W/ 17GA TUOHY, Material Number 332602 Connection device used by physicians to provide various anesthetic and fluid administration devices with a single, common access point to a catheter for delivery of anesthetics. The connector is used in conjunction with the catheter for continuous admi...
The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the catheters from the catheter connectors has been observed.
Terminated
null
2018-07-20
2018-10-10
null
1020
null
3014301563
1
0
Voluntary Action Indicated (VAI)
2018-12-27
0
0
null
false
-1
device_malfunction
connector failure
Z-0453-2021
device
Cardinal Health 200, LLC
Kangaroo Connect 1000ml Bag Set, Non-Sterile, Product Code 77100FD - Product Usage: single-use tubing sets used with pumps in the delivery of nutritional feeding product.
The feeding spike sets may leak at the interface of the tube and spike connector.
Ongoing
null
2020-09-22
2020-11-25
null
197370
null
3034315018
1
0
No Action Indicated (NAI)
2025-09-18
0
0
null
false
-1
device_malfunction
leaking connector interface
Z-1261-2020
device
Siemens Healthcare Diagnostics, Inc.
Atellica CH High Sensitivity C-Reactive Protein (HsCRP) Calibrator - Product Usage: For in vitro diagnostic use in the calibration of ADVIA¿ Chemistry systems for the CardioPhase¿ hsCRP assay
Atellica CH: The Atellica¿ CH High Sensitivity CReactive Protein Calibrator (hsCRP CAL) is for in vitro diagnostic use in calibrating the hsCRP assay using the Atellica¿ CH Analyzer.
Terminated
null
2019-11-19
2020-02-26
null
1306
null
1219913
8
1
Voluntary Action Indicated (VAI)
2022-03-08
1
1
2012-05-10
false
-1
device_malfunction
calibrator defect